UV-Vis Spectrophotometer
LabTech's UV-Vis spectrophotometers provide flexible and efficient operation with excellent stability and repeatability. Large LCD with friendly interface faciliates operation. Both classical and economical, they are suitable for daily routine QA/QC analysis, production analysis, as well as for more specialized applications. Powerful software is included for photometric measurement, spectrum scanning, quantitative determination, and water quality analysis.
Technical Features
-
Real double-beam system.
-
High integration 32Bit CPU guarantees high performance.
-
Five kinds of slits: 0.5nm, 1nm, 2nm, 3nm, and 5nm.
-
Quartz coated and sealed optics prevents corrosion from environment.
-
Light sources automatically record the working time.
-
Automated six-coupled cell enormously shortens the sample replacement time.
-
Intellectual auto-checking function of the software is convenient for trouble shooting and maintenance.
-
Standalone systems are with powerful 32M storage capacity.
-
Options of standalone systems or systems with separate PC controllers.
-
Streamline surface protects sample splashes into internal space.
UV Power- Series
Bandwidth
Optical System
Wavelength Rang
Photometric Method
Wavelength Accuracy
Wavelength Repeatability
Absorbance Range
Photometric Accuracy
Photometric Repeatability
Wavelength Scan Speed
Stray Light
Baseline Straightness
Drift
Noise
Detector
Light Source
Cell Holder
Display
Control Mode
Dimensions (WxDxH) Inch
Weight Lbs
0.5nm / 1nm / 2nm / 3nm / 5nm
Double beam , C-T monochromator, holographic grating
190 ~1100nm
Transmittance, Absorbance, Energy
±0.3nm
±0.1nm
-3 ~ 3Abs
±0.002A (0 ~ 0.5A), ±0.004A (0.5 ~ 1A), ±0.3%T
≤0.001A (0 ~ 0.5A), ≤0.002A (0.5 ~ 1A), ±0.1%T
Adjustable (Fast, Middle, Low)
≤0.03% T
±0.001A
≤0.0002A/h (500nm, 0A)
±0.0002A (500nm, P-P)
Si-Photodiode
Halogen tungsten lamp / Deuterium lamp (Auto-switch)
Automatic six-position cell holder
320 × 240 pixel, LCD
PC control or Local control
20x19x10
49
UV 9100 - Series
Bandwidth
Optical System
Wavelength Rang
Photometric Method
Wavelength Accuracy
Wavelength Repeatability
Absorbance Range
Photometric Accuracy
Photometric Repeatability
Wavelength Scan Speed
Stray Light
Baseline Straightness
Drift
Noise
Detector
Light Source
Cell Holder
Display
Control Mode
Dimensions (WxDxH) Inch
Weight Lbs
0.5nm / 1nm / 2nm / 3nm / 5nm
Double Beam, C-T monochromator, holographic grating
190 ~1100nm
Transmittance, Absorbance, Energy
±0.3nm
±0.1nm
-3 ~ 3Abs
±0.002A (0 ~ 0.5A), ±0.004A (0.5 ~ 1A), ±0.3%T
≤0.001A (0 ~ 0.5A), ≤0.002A (0.5 ~ 1A), ±0.1%T
Adjustable (Fast, Middle, Low)
≤0.05% T
±0.001A
≤0.0003A/h (500nm, 0A)
±0.0002A (500nm, P-P)
Si-Photodiode
Halogen tungsten lamp / Deuterium lamp (Auto-switch)
Automatic six-position cell holder
320 × 240 pixel, LCD
Local control or PC control
20x19x10
49
UV 8100 - Series
Bandwidth
Optical System
Wavelength Rang
Photometric Method
Wavelength Accuracy
Wavelength Repeatability
Absorbance Range
Photometric Accuracy
Photometric Repeatability
Wavelength Scan Speed
Stray Light
Baseline Straightness
Drift
Noise
Detector
Light Source
Cell Holder
Display
Control Mode
Dimensions (WxDxH) Inch
Weight Lbs
0.5nm / 1nm / 2nm / 3nm / 5nm
Split Beam, C-T monochromator, holographic grating
190 ~1100nm
Transmittance, Absorbance, Energy
±0.3nm
±0.2nm
-3 ~ 3Abs
±0.002A (0 ~ 0.5A), ±0.004A (0.5 ~ 1A), ±0.3% T
≤0.001A (0 ~ 0.5A), ≤0.002A (0.5 ~ 1A), ±0.1% T
Adjustable (Fast, Middle, Low)
≤0.05% T
±0.001A
≤0.0005A/h (500nm, 0A)
±0.0002A (500nm, P-P)
Si-Photodiode
Halogen tungsten lamp / Deuterium lamp (Auto-switch)
Automatic six-position cell holder
320 × 240 pixel, LCD
Local control or PC control
20x19x10
49
Bluestar - Series
Bandwidth
Optical System
Wavelength Rang
Photometric Method
Wavelength Accuracy
Wavelength Repeatability
Absorbance Range
Photometric Accuracy
Photometric Repeatability
Wavelength Scan Speed
Stray Light
Baseline Straightness
Drift
Noise
Detector
Light Source
Cell Holder
Display
Control Mode
Dimensions (WxDxH) Inch
Weight Lbs
1nm / 2nm
Split Beam, C-T monochromator, holographic grating
190 ~1100nm
Transmittance, Absorbance, Energy
±0.8nm
±0.2nm
-3 ~ 3Abs
±0.002A (0 ~ 0.5A), ±0.004A (0.5 ~ 1A), ±0.3% T
≤0.001A (0 ~ 0.5A), ≤0.002A (0.5 ~ 1A), ±0.1% T
Adjustable (Fast, Middle, Low)
≤0.05% T
±0.002A
≤0.0005A/h (500nm, 0A)
±0.0002A (500nm, P-P)
Si-Photodiode
Halogen tungsten lamp / Deuterium lamp (Auto-switch)
Automatic six-position cell holder
320 × 240 pixel, LCD
Local control or PC control
20x19x10
49